T1 - Ultrasensitive non-enzymatic electrochemical detection of paraoxon-ethyl in fruit samples using 2D Ti3C2Tx/MWCNT-OH
AU - Aris, Asmi
AU - Wahyuni, Wulan Tri
AU - Putra, Budi Riza
AU - Hermawan, Angga
AU - Nugroho, Ferry Anggoro Ardy
AU - Seh, Zhi Wei
AU - Khalil, Munawar
N1 - Publisher Copyright:
© The Royal Society of Chemistry 2025.
PY - 2025/1/3
Y1 - 2025/1/3
N2 - This study reports on the development of a highly sensitive non-enzymatic electrochemical sensor based on a two-dimensional Ti3C2Tx/MWCNT-OH nanocomposite for the detection of paraoxon-based pesticide. The synergistic effect between the Ti3C2Tx nanosheet and the functionalized multi-walled carbon nanotubes enhanced the sensor’s conductivity and catalytic activity. The nanocomposite demonstrates superior electrochemical and electroanalytical performance compared to the pristine Ti3C2Tx and MWCNT-OH in detecting paraoxon-ethyl in fruit samples (green and red grapes), with a linear response range from 0.1 to 100 μM, a low limit of detection (LOD) of 10 nM, limit of quantitation (LOQ) of 70 nM, and sensitivity of 0.957 µA μM−1 cm−2 at pH 8. Furthermore, the sensors maintain excellent selectivity and effectiveness in detecting paraoxon-ethyl even in the presence of various interferents, including diazinon, carbaryl, Fe2+, NO2−, NO3−, ascorbic acid, and glucose. The facile fabrication and enhanced sensing capabilities of the Ti3C2Tx/MWCNT-OH nanocomposite position it as a reliable, cost-effective, and sustainable alternative to conventional detection systems for monitoring pesticide residues in agricultural products.
AB - This study reports on the development of a highly sensitive non-enzymatic electrochemical sensor based on a two-dimensional Ti3C2Tx/MWCNT-OH nanocomposite for the detection of paraoxon-based pesticide. The synergistic effect between the Ti3C2Tx nanosheet and the functionalized multi-walled carbon nanotubes enhanced the sensor’s conductivity and catalytic activity. The nanocomposite demonstrates superior electrochemical and electroanalytical performance compared to the pristine Ti3C2Tx and MWCNT-OH in detecting paraoxon-ethyl in fruit samples (green and red grapes), with a linear response range from 0.1 to 100 μM, a low limit of detection (LOD) of 10 nM, limit of quantitation (LOQ) of 70 nM, and sensitivity of 0.957 µA μM−1 cm−2 at pH 8. Furthermore, the sensors maintain excellent selectivity and effectiveness in detecting paraoxon-ethyl even in the presence of various interferents, including diazinon, carbaryl, Fe2+, NO2−, NO3−, ascorbic acid, and glucose. The facile fabrication and enhanced sensing capabilities of the Ti3C2Tx/MWCNT-OH nanocomposite position it as a reliable, cost-effective, and sustainable alternative to conventional detection systems for monitoring pesticide residues in agricultural products.
UR - http://www.scopus.com/inward/record.url?scp=85215839833&partnerID=8YFLogxK
U2 - 10.1039/d4nr04060k
DO - 10.1039/d4nr04060k
M3 - Article
C2 - 39836463
AN - SCOPUS:85215839833
SN - 2040-3364
VL - 17
SP - 2554
EP - 2566
JO - Nanoscale
JF - Nanoscale
IS - 5
ER -