T1 - The effect of anticoagulant types on the in vitro analysis of clopidogrel in human plasma using liquid chromatography tandem-mass spectrometry
AU - Harahap, Yahdiana
AU - Maulidina, Anisa
AU - Ramadon, Delly
N1 - Publisher Copyright:
© 2018 The Authors.
PY - 2018/12/1
Y1 - 2018/12/1
N2 - Objective: The aim of this study was to optimize and validate a plasma clopidogrel analysis method using liquid chromatography tandem-mass spectrometry. Methods: Plasma samples were analyzed using a BEH C18 column (1.7 µ 100 mm×2.1 mm), the mobile phase was 0.1% formic acid in acetonitrile (30:70, v/v). The flow rate was 0.2 mL/min, with a column temperature set to 35°C, an injection volume of 5 µL, an analysis time of 4 min, and irbesartan as the internal standard. Aliquots were obtained by liquid-liquid extraction using ammonium acetate and diethyl ether. The stability and peak area ratio of the respective plasma area responses were evaluated using ANOVA. Results: No significant differences (p>0.05) were observed between anticoagulants regarding analyte stability. However, the peak area ratio showed significant differences (p<0.05) between the anticoagulants. The accuracy and precision of the analysis with citrate, heparin, and ethylenediaminetetraacetic acid (EDTA) plasma met the quality requirements, and a linear calibration curve was created with concentrations ranging from 0.02 to 5.0 ng/mL. Conclusion: The results showed that improved analysis of clopidogrel was achieved using citrate or heparin plasma compared with EDTA plasma.
AB - Objective: The aim of this study was to optimize and validate a plasma clopidogrel analysis method using liquid chromatography tandem-mass spectrometry. Methods: Plasma samples were analyzed using a BEH C18 column (1.7 µ 100 mm×2.1 mm), the mobile phase was 0.1% formic acid in acetonitrile (30:70, v/v). The flow rate was 0.2 mL/min, with a column temperature set to 35°C, an injection volume of 5 µL, an analysis time of 4 min, and irbesartan as the internal standard. Aliquots were obtained by liquid-liquid extraction using ammonium acetate and diethyl ether. The stability and peak area ratio of the respective plasma area responses were evaluated using ANOVA. Results: No significant differences (p>0.05) were observed between anticoagulants regarding analyte stability. However, the peak area ratio showed significant differences (p<0.05) between the anticoagulants. The accuracy and precision of the analysis with citrate, heparin, and ethylenediaminetetraacetic acid (EDTA) plasma met the quality requirements, and a linear calibration curve was created with concentrations ranging from 0.02 to 5.0 ng/mL. Conclusion: The results showed that improved analysis of clopidogrel was achieved using citrate or heparin plasma compared with EDTA plasma.
KW - Citrate
KW - Clopidogrel
KW - Ethylenediaminetetraacetic acid
KW - Heparin
KW - Liquid chromatography tandem-mass spectrometry
UR - http://www.scopus.com/inward/record.url?scp=85071868193&partnerID=8YFLogxK
U2 - 10.22159/ijap.2018.v10s1.87
DO - 10.22159/ijap.2018.v10s1.87
M3 - Article
AN - SCOPUS:85071868193
SN - 0975-7058
VL - 10
SP - 392
EP - 395
JO - International Journal of Applied Pharmaceutics
JF - International Journal of Applied Pharmaceutics
IS - Special Issue 1
ER -