The characteristics of small field beam quality and output factor of 6 MV FFF

A. Azzi, D. Ryangga, S. A. Pawiro

Research output: Contribution to journalConference article

Abstract

Flattening filter free linear accelerator (FFF LINAC) has been installed in Indonesia. To ensure the accuracy of FFF irradiation, we evaluate the characteristic of FFF in regular and small fields. We employed the Monte Carlo (MC) simulation of FFF LINAC 6 MV as a standard reference in this study. Then, we compared the calculation of FFF beam at central and lateral axis to the measurement and TPS at 10 × 10, 1 × 1, 2 × 2, 3 × 3, 4 × 4 cm2 field sizes. Output factor (OF) and beam quality such that TPR20,10 and penumbra of FFF LINAC were evaluated. TPR20,10(S) for 10 × 10 cm2 was 0.01 differ between MC and measurement whereas it varied from 0.588 to 0.703 for small fields. Dose difference of lateral axis was agreed within 3% except for the penumbra region. Output factor of small fields measurement indicated that field size of 1 × 1 cm2 had a large discrepancy to MC according to this works. The TPR20,10(10) of MC, TPS, and measurement was not a significant difference, while the OF was tending to a large deviation in 1 × 1 cm2. The results showed that our FFF LINAC could be used for small field until 2 × 2 cm2.

Original languageEnglish
Article number012056
JournalJournal of Physics: Conference Series
Volume1248
Issue number1
DOIs
Publication statusPublished - 6 Jun 2019
Event18th Asia-Oceania Congress of Medical Physics, AOCMP 2018 and 16th South-East Asia Congress of Medical Physics, SEACOMP 2018 - Bangsar South, Kuala Lumpur, Malaysia
Duration: 11 Nov 201814 Nov 2018

Fingerprint Dive into the research topics of 'The characteristics of small field beam quality and output factor of 6 MV FFF'. Together they form a unique fingerprint.

  • Cite this