Abstract
To answer the need for traceability in micro flow area for medical industry, Flow Laboratory of Metrology LIPI in cooperation with Physics Department of University of Indonesia, developed a compact standard based on piston displacement concept. The standard consists of glass syringe and linear actuator controlled by Field Programmable Gate Array (FPGA). The volume generation uncertainty is 0.6% to 0.7% at 6.5 mL to 13.4 mL. Flow-rate uncertainty is 1.6% to 11.6% at 90.3 mL/hr to 1177.1 mL/hr.
Original language | English |
---|---|
Publication status | Published - 2015 |
Event | 21st IMEKO World Congress on Measurement in Research and Industry - Prague, Czech Republic Duration: 30 Aug 2015 → 4 Sept 2015 |
Conference
Conference | 21st IMEKO World Congress on Measurement in Research and Industry |
---|---|
Country/Territory | Czech Republic |
City | Prague |
Period | 30/08/15 → 4/09/15 |
Keywords
- FPGA
- Micro-flow
- Syringe pump