Capacitance-digital and impedance converter as electrical tomography measurement system for biological tissue

Mila Izzatul Ikhsanti, Rana Bouzida, Sastra Kusuma Wijaya, Rohmadi, Imamul Muttakin, Warsito P. Taruno

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

2 Citations (Scopus)

Abstract

This research aims to explore the feasibility of capacitance-digital converter and impedance converter for measurement module in electrical capacitance tomography (ECT) system. ECT sensor used was a cylindrical sensor having 8 electrodes. Absolute capacitance measurement system based on Sigma Delta Capacitance-to-Digital-Converter AD7746 has been shown to produce measurement with high resolution. Whereas, capacitance measurement with wide range of frequency is possible using Impedance Converter AD5933. Comparison of measurement accuracy by both AD7746 and AD5933 with reference of LCR meter was evaluated. Biological matters represented in water and oil were treated as object reconstructed into image using linear back projection (LBP) algorithm.

Original languageEnglish
Title of host publicationBiomedical Engineering's Recent Progress in Biomaterials, Drugs Development, and Medical Devices
Subtitle of host publicationProceedings of the 1st International Symposium of Biomedical Engineering, ISBE 2016
EditorsYudan Whulanza, Sugeng Supriadi, Muhamad Sahlan, Basari
PublisherAmerican Institute of Physics Inc.
ISBN (Electronic)9780735414853
DOIs
Publication statusPublished - 21 Feb 2017
Event1st International Symposium of Biomedical Engineering: Biomedical Engineering's Recent Progress in Biomaterials, Drugs Development, and Medical Devices, ISBE 2016 - Depok City, Indonesia
Duration: 31 May 20161 Jun 2016

Publication series

NameAIP Conference Proceedings
Volume1817
ISSN (Print)0094-243X
ISSN (Electronic)1551-7616

Conference

Conference1st International Symposium of Biomedical Engineering: Biomedical Engineering's Recent Progress in Biomaterials, Drugs Development, and Medical Devices, ISBE 2016
Country/TerritoryIndonesia
CityDepok City
Period31/05/161/06/16

Fingerprint

Dive into the research topics of 'Capacitance-digital and impedance converter as electrical tomography measurement system for biological tissue'. Together they form a unique fingerprint.

Cite this