Abstract

The use of air purifier to clean the room with poor ventilation is considered as a fast and common solution to minimize the spread of SARS-CoV-2 nowadays. However, the efficacy of the filter within the air purifier must be tested priorly to ensure the cleanliness of the air, since the size of this novel virus is very small. Inexpensive particulate counter which is common in the market, such as HT-9600 particle counter can be used to test the efficacy of the filter if the device is properly calibrated. This paper aims to calibrate the HTI portable particulate counter to Solair3100 which has been calibrated by NIST using ISO 21501-4 and ISO 17025 standards. Both devices were used to measure the amount of particle in various places to see their characteristic. The same nonlinear trend produced by both devices in reading the amount of particle in different places made the calibration effort can be done. Data used for calibration were the average reading of the number of particles from various conditions. Calibration curve of HT-9600 particle counter was obtained by plotting the average readings of both devices in a single plot. The obtained calibration curve is a non-linear plot which represents the characteristic of particle counter in reading the amount of particle based on its respective size. The obtained calibration function from the curve shows great tendency in reading the right amount of particle in the air with minimum error.

Original languageEnglish
Title of host publicationAIP Conference Proceedings
EditorsAndyka Kusuma, Jaka Fajar Fatriansyah, Radon Dhelika, Mochamad Adhiraga Pratama, Ridho Irwansyah, Imam Jauhari Maknun, Wahyuaji Narottama Putra, Romadhani Ardi, Ruki Harwahyu, Yulia Nurliani Harahap, Kenny Lischer
PublisherAmerican Institute of Physics Inc.
Edition1
ISBN (Electronic)9780735446410
DOIs
Publication statusPublished - 6 Feb 2024
Event17th International Conference on Quality in Research, QiR 2021 in conjunction with the International Tropical Renewable Energy Conference 2021, I-Trec 2021 and the 2nd AUN-SCUD International Conference, CAIC-SIUD - Virtual, Online, India
Duration: 13 Oct 202115 Oct 2021

Publication series

NameAIP Conference Proceedings
Number1
Volume2710
ISSN (Print)0094-243X
ISSN (Electronic)1551-7616

Conference

Conference17th International Conference on Quality in Research, QiR 2021 in conjunction with the International Tropical Renewable Energy Conference 2021, I-Trec 2021 and the 2nd AUN-SCUD International Conference, CAIC-SIUD
Country/TerritoryIndia
CityVirtual, Online
Period13/10/2115/10/21

Fingerprint

Dive into the research topics of 'Calibration of low-cost portable particulate counter (PC) to standardized-certified PC'. Together they form a unique fingerprint.

Cite this